Technical Information

Proline Promass 80A, 83A

Coriolis Mass Flow Measuring System
The single-tube system for highly accurate measurement of 
very small flows

Application

The Coriolis measuring principle operates independently 
of the physical fluid properties, such as viscosity and 
density.

• Suitable for continuous measurement, filling and 

dosing of very small flows.

• Extremely accurate measurement of liquids and 

gases such as emulsions, additives, flavouring, 
insulin, gases for high pressure and low pressure

• Fluid temperatures up to +200 °C (+392 °F)
• Process pressures up to 400 bar (5800 psi)

Approvals for hazardous area:
• ATEX, FM, CSA, TIIS, IECEx, NEPSI

Approvals in the food industry/hygiene sector:
• 3A, FDA, EHEDG

Connection to process control system:
• HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, 

MODBUS

Relevant safety aspects:
• Pressure Equipment Directive, SIL-2
• Purge connection or rupture disk (optional)

Your benefits

The Promass measuring devices make it possible to 
simultaneously record several process variables (mass/
density/temperature) for various process conditions 
during measuring operation.

The 

Proline transmitter concept

 comprises:

• Modular device and operating concept resulting in a 

higher degree of efficiency

• Software options for batching and concentration 

measurement for extended range of application

• Diagnostic ability and data back-up for increased 

process quality

The 

Promass sensors

, tried and tested in over 

100000 applications, offer:
• Multivariable flow measurement in compact design
• Insensitivity to vibrations thanks to balanced single-

tube measuring system

• Immune from external piping forces due to robust 

design

• Easy installation without taking inlet and outlet runs 

into consideration

Summary of Contents for Proline Promass 80A

Page 1: ...ystem HART PROFIBUS DP PA FOUNDATION Fieldbus MODBUS Relevant safety aspects Pressure Equipment Directive SIL 2 Purge connection or rupture disk optional Your benefits The Promass measuring devices ma...

Page 2: ...Inlet and outlet run 19 Length of connecting cable 19 System pressure 19 Operating conditions Environment 20 Ambient temperature range 20 Storage temperature 20 Ambient class 20 Degree of protection...

Page 3: ...he inlet of the tubes 2 and acceleration at the outlet 3 a0003383 The phase difference A B increases with increasing mass flow Electrodynamic sensors register the tube oscillations at the inlet and ou...

Page 4: ...ions A a0003679 Single tube system for highly accurate measurement of very small flows Nominal diameters DN 1 to 4 1 24 to 1 8 Material Stainless steel EN 1 4539 ASTM 904L EN 1 4404 ASTM 316L process...

Page 5: ...ments for the secondary containment and fluid temperatures up to 150 C 302 F Nominal diameters DN 8 to 80 3 8 to 3 Material Titanium Ti Grade 2 Ti Grade 9 Documentation No TI102D P a0006828 Single ben...

Page 6: ...es See information in the Limiting flow Section 21 Operable flow range Greater than 1000 1 Flow rates above the preset full scale value do not overload the amplifier i e the totalizer values are regis...

Page 7: ...ull scale frequency 2 to 1000 Hz fmax 1250 Hz on off ratio 1 1 pulse width max 2 s Pulse output pulse value and pulse polarity can be selected pulse width adjustable 0 5 to 2000 ms PROFIBUS PA interfa...

Page 8: ...djustment Measuring mode Totalizer control Bus address can be set at the measuring device via miniature switches or the on site display optional Available output combination 11 PROFIBUS PA interface P...

Page 9: ...Link Master function LM is supported Signal on alarm Current output Failsafe mode selectable e g in accordance with NAMUR Recommendation NE 43 Pulse frequency output Failsafe mode selectable Status ou...

Page 10: ...MODBUS RS485 B PA FF with reverse polarity protection Terminal No 27 DP A PA FF MODBUS RS485 A PA FF with reverse polarity protection e Earth terminal signal cable screen fieldbus cable RS485 line f S...

Page 11: ...Terminal No inputs outputs Order variant 20 21 22 23 24 25 26 27 Fixed communication boards permanent assignment 83 A Frequency output Current output HART 83 B Relay output Relay output Frequency out...

Page 12: ...DBUS RS485 83 P Current output Frequency output Status input PROFIBUS DP 83 V Relay output 2 Relay output 1 Status input PROFIBUS DP 83 W Relay output Current output 3 Current output 2 Current output...

Page 13: ...IEC EN 61326 and NAMUR recommendation NE 21 43 Power consumption AC 15 VA including sensor DC 15 W including sensor Switch on current Max 13 5 A 50 ms at 24 V DC Max 3 A 5 ms at 260 V AC Power supply...

Page 14: ...w liquid Promass 83A 0 10 o r Promass 80A 0 15 o r Mass flow gas Promass 83A 80A 0 50 o r Density liquid 0 0005 g cc under reference conditions 0 0005 g cc after field density calibration under proces...

Page 15: ...ow and volume flow liquids Promass 83A 80A 0 05 o r Mass flow gases Promass 83A 80A 0 25 o r Density liquids 0 00025 g cc 1 g cc 1 kg l Temperature 0 25 C 0 0025 T C 1 F 0 003 T 32 F T medium temperat...

Page 16: ...calibration pressure and the process pressure does not have any effect on the accuracy Design fundamentals Dependent on the flow Flow Zero point stability Base accuracy 100 Max measured error Base ac...

Page 17: ...pipe support is recommended for heavy sensors Mounting location Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors For this reason avoid the following mou...

Page 18: ...the transmitter housing is above or below the pipe This means that no gas bubbles or solids deposits can form in the bent measuring tube single tube system The sensor cannot be installed suspended i e...

Page 19: ...n be ordered as accessories from Endress Hauser are available for the sensors Zero point adjustment All Promass devices are calibrated to state of the art technology The zero point determined in this...

Page 20: ...accordance with IEC 68 2 31 Vibration resistance Acceleration up to 1g 10 to 150 Hz following IEC 68 2 6 CIP cleaning yes SIP cleaning yes Electromagnetic compatibility EMC To IEC EN 61326 and NAMUR...

Page 21: ...detection Rupture disk optional Further informationen 35 Limiting flow See information in the Measuring range Section 6 Select nominal diameter by optimising between required flow range and permissib...

Page 22: ...e 2300 a0003379 p pressure loss mbar v kinematic viscosity m2 s g mass flow kg s density kg m d inside diameter of measuring tubes m K K1 constant depends on nominal diameter 1 To compute the pressure...

Page 23: ...al construction Design dimensions Dimensions Field housing compact version powder coated die cast aluminum 24 Transmitter compact version stainless steel 25 Transmitter remote version connection housi...

Page 24: ...lay 1 depends on the process connection Dimensions for the holes e on the baseplate for a table a wall or post mounting Dimensions G x J All dimensions in mm DN A A B C D E F G H J K L M N O U di 1 8...

Page 25: ...mensions in mm A A B B C D E F G H J K L M 10 4 9 53 9 45 8 54 8 11 7 32 7 01 0 34 M8 3 94 5 12 3 94 5 67 6 69 13 7 Blind version without display All dimensions in inch A B C Esc E Nicht unter Spannun...

Page 26: ...45 50 81 53 L M N O P Q R S T 1 95 53 102 81 5 11 5 192 8 M5 20 2 x 6 5 1 Fixing bolt for wall assembly M6 screw head max 10 5 mm All dimensions in mm A B C D E F G H J K 8 46 9 84 3 56 6 27 5 31 3 5...

Page 27: ...sion connetion housing a0002516 Dimensions in SI units Dimensions in US units DN A B C 1 118 5 137 5 120 2 118 5 137 5 120 4 118 5 137 5 130 All dimensions in mm DN A B C 1 24 4 67 5 41 4 72 1 12 4 67...

Page 28: ...2 2 372 AF 11 16 1 4 4 1 497 AF 11 16 3 5 4 2 497 AF 11 16 3 0 1 3A version can be supplied Ra 0 4 m 240 grit Only for 1 4539 904L 2 High pressure version All dimensions in mm Tri Clamp connection 3A...

Page 29: ...393 95 2 15 7 4 15 7 20 7 66 5 1 1 2 Cl 150 475 88 9 15 7 4 15 7 17 7 60 5 1 8 2 Cl 300 475 95 2 15 7 4 15 7 20 7 66 5 1 8 4 Cl 150 600 88 9 15 7 4 15 7 17 7 60 5 3 5 4 Cl 300 600 95 2 15 7 4 15 7 20...

Page 30: ...68 AF NPT 3 5 41 568 AF NPT 3 0 High pressure version only available as 1 4539 904L All dimensions in mm Further dimensions Mounting kit SWAGELOK connection 1 4539 904L DN L P Q di 1 359 6 AF 7 16 1 8...

Page 31: ...0 06 1 8 1 19 6 AF 11 16 0 14 1 8 2 19 6 AF 11 16 0 12 1 3A version can be supplied Ra 0 4 m 240 grit Only for 1 4539 904L 2 High pressure version All dimensions in inch Tri Clamp connection 3A versio...

Page 32: ...6 4 0 6 0 56 2 8 0 04 1 12 10K 19 3 8 0 6 4 0 6 1 12 2 8 0 07 1 12 20K 19 3 8 0 6 4 0 6 0 56 2 8 0 07 1 8 10K 24 3 8 0 6 4 0 6 1 12 2 8 0 14 1 8 20K 24 3 8 0 6 4 0 6 0 56 2 8 0 14 Loose flanges not w...

Page 33: ...3 4 1 4 NPT 0 14 1 8 1 22 4 AF 3 4 1 4 NPT 0 12 1 High pressure version only available as 1 4539 904L All dimensions in inch Mounting kit SWAGELOK connection 1 4539 904L DN L P Q di 1 24 14 2 AF 7 16...

Page 34: ...y with a dry inert gas Use only low gauge pressure to purge Maximum pressure 5 bar 72 5 psi a0003187 Dimensions in SI units Dimensions in US units DN A B C G L R 1 70 0 77 0 33 0 NPT 178 47 0 2 70 0 7...

Page 35: ...ntainment function if a rupture disk is used It is not permitted to open the connections or remove the rupture disk Caution The existing connection nozzles are not designed for a rinse or pressure mon...

Page 36: ...kg 11 lbs Weight in SI units Weight in US units DN mm 1 2 4 Compact version 10 11 15 Remote version 8 9 13 All values weight refer to devices with EN DIN PN 40 flanges Weight information in kg DN inch...

Page 37: ...remote version Stainless steel 1 4301 304 Process connections Measuring tube s Stainless steel 1 4539 904L Alloy C 22 2 4602 N 06022 Seals Welded process connections without internal seals Process con...

Page 38: ...loy C 22 Loose flanges not wetted 1 4404 316L a0003312 Tri Clamp process connection The Clamp connections are suited up to a maximum pressure of 16 bar 232 psi Please observe the operating limits of t...

Page 39: ...crewed mounting kit 1 4401 316 a0003310 Process connections Welded process connections 4 VCO 4 coupling 1 2 Tri Clamp Screwed on process connections flanges EN 1092 1 DIN 2501 ASME JIS 1 4 NPT threade...

Page 40: ...operation with three optical keys S O F Application specific Quick Setup menus for straightforward commissioning Language group Language groups available for operation in different countries Western E...

Page 41: ...rganization The device thus meets all the requirements of the following specifications Certified to PROFIBUS Profile Version 3 0 device certification number on request The device can also be operated...

Page 42: ...3F TI101D Promass 80H 83H TI074D Promass 80I 83I TI075D Promass 80M 83M TI102D Promass 80P 83P TI078D Promass 80S 83S TI076D Operating Instructions Description of Device Functions Promass 80 HART BA05...

Page 43: ...elok Co Solon USA HART Registered trademark of HART Communication Foundation Austin USA PROFIBUS Registered trademark of the PROFIBUS User Organisation Karlsruhe Germany FOUNDATION Fieldbus Registered...

Page 44: ...national Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 715 81 00 Fax 41 61 715 25 00 www endress com info ii endress com TI054D 06 en 10 09 71104043 FM...