Commissioning
Proline Promag W 400 HART
88
Hauser
10.4.4 Configuring the current output
The
Current output
wizard guides you systematically through all the parameters that have
to be set for configuring the current output.
Navigation
"Setup" menu → Current output 1
‣
Current output 1
Assign current output 1
Current span
0/4 mA value
20 mA value
Fixed current
Failure mode
Failure current
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign current output
–
Select process variable for
current output.
• Volume flow
• Mass flow
• Corrected volume
flow
• Flow velocity
• Conductivity
• Corrected
conductivity
• Temperature
• Electronic
temperature
Volume flow
Current span
–
Select current range for
process value output and
upper/lower level for alarm
signal.
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
• Fixed current
Country-specific:
• 4...20 mA NAMUR
• 4...20 mA US
0/4 mA value
One of the following options is
selected in the
Current span
parameter (→ 88):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Enter 4 mA value.
Signed floating-point
number
Country-specific:
• 0 l/h
• 0 gal/min (us)