Installation
Proline Promag W 400 HART
36
Hauser
NOTICE
Plug of the smart sensor electronics module connected incorrectly!
No measuring signal is output.
‣
Plug in the plug of the smart sensor electronics module as per the coding.
PUSH
TO
REMOVE
A0021585
‣
Reverse the procedure to reassemble the measuring device.
6.2.6
Turning the display module
The display module can be turned to optimize display readability and operability.
TX 20
4 x
1.
2.
PUSH
TO
REMOVE
PUSH
TO
REMOVE
3.
4.
A0032091
1. Loosen the fixing screws of the housing cover (when reassembling, pay attention to
the tightening torque → 36).
2. Open the housing cover.
3. Unlock the display module.
4. Pull out the display module and turn it to the desired position in increments of 90°.
Reassembling the transmitter housing
L
WARNING
Excessive tightening torque applied to the fixing screws!
Risk of damaging the plastic transmitter.
‣
Tighten the fixing screws as per the tightening torque:
Step
(see graphic)
Fixing screw
Tightening torque for housing made of:
Aluminum
Plastic
1
Housing cover
2.5 Nm (1.8 lbf ft)
1 Nm (0.7 lbf ft)
‣
Reverse the procedure to reassemble the measuring device.