Proline Promag W 400 HART
Installation
Hauser
27
3. If using ground disks, comply with the Installation Instructions provided.
4. Observe required screw tightening torques → 27.
5. Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0029263
Mounting the seals
L
CAUTION
An electrically conductive layer could form on the inside of the measuring tube!
Risk of measuring signal short circuit.
‣
Do not use electrically conductive sealing compounds such as graphite.
Comply with the following instructions when installing seals:
1. Make sure that the seals do not protrude into the piping cross-section.
2. For DIN flanges: only use seals according to DIN EN 1514-1.
3. For "hard rubber" lining: additional seals are
always
required.
4. For "polyurethane" lining: generally additional seals are
not
required.
Mounting the ground cable/ground disks
Comply with the information on potential equalization and detailed mounting instructions
for the use of ground cables/ground disks .
Screw tightening torques
Please note the following:
• The screw tightening torques listed below apply only to lubricated threads and to pipes
not subjected to tensile stress.
• Tighten the screws uniformly and in diagonally opposite sequence.
• Overtightening the screws will deform the sealing faces or damage the seals.
Screw tightening torques for EN 1092-1 (DIN 2501), PN 6/10/16/25/40
Nominal diameter
Pressure rating
Threaded fasteners
Max. screw tightening torque [Nm]
[mm]
[bar]
[mm]
Hard rubber
Polyurethane
25
PN 40
4 × M12
–
15
32
PN 40
4 × M16
–
24
40
PN 40
4 × M16
–
31
50
PN 40
4 × M16
48
40
65
PN 16
8 × M16
32
27
65
PN 40
8 × M16
32
27
80
PN 16
8 × M16
40
34
80
PN 40
8 × M16
40
34
100
PN 16
8 × M16
43