Proline Promag W 400 HART
Commissioning
Hauser
111
Frequency output simulation 1 to n
Frequency value 1 to n
Pulse output simulation 1 to n
Pulse value 1 to n
Switch output simulation 1 to n
Switch status 1 to n
Device alarm simulation
Diagnostic event category
Diagnostic event simulation
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign simulation process variable
–
Select a process variable for
the simulation process that is
activated.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Flow velocity
• Conductivity
• Corrected
conductivity
• Temperature
Off
Process variable value
One of the following options is
selected in the
Assign
simulation process variable
parameter (→ 111):
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Corrected conductivity
• Temperature
Enter the simulation value for
the selected process variable.
Depends on the
process variable
selected
0
Status input simulation
For the following order code:
• "Output; input", option I
"4-20mA HART, 2x pul./
freq./switch output; status
input"
• "Output; input", option J
"4-20mA HART, certified
pulse output, switch output;
status input"
Switch simulation of the status
input on and off.
• Off
• On
Off
Input signal level
In the
Status input simulation
parameter, the
On
option is
selected.
Select the signal level for the
simulation of the status input.
• High
• Low
High
Current output 1 simulation
–
Switch the simulation of the
current output on and off.
• Off
• On
Off