5 Block USER INTERFACE
Device Functions Proline Promag 55 FOUNDATION Fieldbus
30
Hauser
FORMAT
(2422)
!
Note!
This function is not available unless a number was selected in the ASSIGN function
(2420).
Use this function to define the maximum number of places after the decimal point for the
second value displayed in the additional line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
!
Note!
• Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device cannot always
be displayed, depending on this setting and the engineering unit. In these instances an
arrow appears on the display between the measured value and the engineering unit
(e.g. 1.2
→
kg/h), indicating that the measuring system is computing with more
decimal places than can be shown on the display.
DISPLAY MODE
(2423)
!
Note!
This function is only available if VOLUME FLOW BARGRAPH IN % or MASS FLOW
BARGRAPH IN % was selected in the function ASSIGN (2420).
Use this function to define the format of the bar graph.
Options:
STANDARD (Simple bar graph with 25 / 50 / 75% gradations and integrated sign).
A0001258
SYMMETRY (Symmetrical bar graph for positive and negative directions of flow, with
–50 / 0 / +50% gradations and integrated sign).
A0001259
Factory setting:
STANDARD
Function description
USER INTERFACE
→
ADDITIONAL LINE
→