Device Functions Proline Promag 55 FOUNDATION Fieldbus
5 Block USER INTERFACE
Hauser
25
FORMAT
(2202)
Use this function to define the maximum number of places after the decimal point
displayed for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
!
Note!
• Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device cannot always
be displayed, depending on this setting and the engineering unit. In these instances an
arrow appears on the display between the measured value and the engineering unit
(e.g. 1.2
→
kg/h), indicating that the measuring system is computing with more
decimal places than can be shown on the display.
Function description
USER INTERFACE
→
MAIN LINE
→
CONFIGURATION