Flow damping
Navigation
Expert → Sensor → Process param. → Flow damping (6661)
Description
Use this function to enter flow damping. Reduction of the variability of the flow measured
value (in relation to interference). For this purpose, the depth of the flow filter is adjusted:
when the filter setting increases, the reaction time of the device also increases.
User entry
0 to 15
Factory setting
4
Additional information
User entry
• Value = 0: no damping
• Value > 0: damping is increased
• 0 is a weak damping and 15 a strong one.
• A damping of 0 is not recommended, as the measuring signal is then so noisy that it
is almost impossible to carry out a measurement.
• The damping depends on the measuring period and the filter type selected.
• An increase or decrease in the damping depends on the application.
Effect
The damping affects the following variables of the device:
Flow override
Navigation
Expert → Sensor → Process param. → Flow override (1839)
Description
Use this function to select whether to interrupt the evaluation of measured values. This is
useful for the cleaning processes of a pipeline, for example.
Selection
• Off
• On
Factory setting
Off
Additional information
Description
Flow override is active
• The diagnostic message diagnostic message
C453 Flow override
is displayed.
• Output values
– Temperature: proceeding output
– Totalizers 1-3: Stop being totalized
Positive zero return can also be enabled via the Status input:
Assign stat.inp.