Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
49
Description
Displays the current totalizer overflow.
User interface
Integer with sign
Additional information
Description
If the current totalizer reading exceeds 7 digits, which is the maximum value range that
can be displayed by the operating tool, the value above this range is output as an overflow.
The current totalizer value is therefore the sum of the overflow value and the totalizer
value from the
Totalizer val. 1 to n
parameter.
User interface
The unit of the selected process variable is specified for the totalizer in the
Unit
totalizer
parameter (→ 168).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the
Totalizer val. 1
parameter: 1
968
457 m
3
• Value in the
Tot. overflow 1
parameter: 2 ⋅ 10
7
(2 overflows) = 20
000
000 [m
3
]
• Current totalizer reading: 21
968
457 m
3
"Input values" submenu
Navigation
Expert → Sensor → Measured val. → Input values
‣
Input values
‣
Current input 1 to n
‣
Val.stat.inp. 1 to n
"Current input 1 to n" submenu
Navigation
Expert → Sensor → Measured val. → Input values → Current
input 1 to n
‣
Current input 1 to n
Measured val. 1 to n (1603–1 to n)
Measur. curr. 1 to n (1604–1 to n)