MMP20 SONO Mobil HD2 (option E)
Page 32
Hauser
Power supply:
7V to 24V-DC
Current consumption:
100mA @ 12V/DC during 2-3 sec. measuring cycle time
Measuring range:
0 to 40% vol. water content
Accuracy (in % vol. water content):
±0.2%
Conductivity range:
0 to 5dS/m
Repeatability:
±0.3%
Temperature drift:
±0.3%
Material temperature measuring
range:
-15°C to 50°C
Material temperature accuracy:
±0.5°C (when fully and permanently in material)
Measuring volume:
1.0L
≅
130x100mm diameter
Probe temperature range:
-15°C to 50°C
Calibration:
Pre-programmed calibrations for sand, gravel and grit
Own calibrations possible,
Memory capacity for up to 15
calibration curves,
Calibration curve
possible
for the dielectric constant
Probe degree of protection:
Water-tight potted, IP68 protection
Probe dimensions:
155 x Ø63mm
Rod length:
130mm
Rod diameter:
6mm
Interfaces:
1.5m cable with 7-pin coupling socket