Micropilot FMR62B HART
Installation
Hauser
15
The maximum length of the nozzle H
max
depends on the nozzle diameter D
D
H
max
H
øD
max
80 to 100 mm (3.2 to 4 in)
1
750 mm (70 in)
100 to 150 mm (4 to 6 in)
2
200 mm (88 in)
≥ 150 mm (6 in)
3
300 mm (132 in)
In the case of longer nozzles, reduced measuring performance must be expected.
Please note the following:
• The end of the nozzle must be smooth and free from burrs.
• The edge of the nozzle should be rounded.
• Mapping must be performed.
• Please contact the manufacturer's support department for applications with nozzles
that are higher than indicated in the table.
10.3.5 Turning the housing
The housing can be rotated up to 380° by loosening the locking screw.
Your benefits
• Easy installation due to optimum alignment of housing
• Easily accessible device operation
• Optimum readability of the local display (optional)
1
1
1
B
C
D
A
A0046660
A
Plastic single compartment housing (no locking screw)
B
Aluminum single compartment housing
C
Dual compartment housing
D
Dual compartment housing, L-form
1
Locking screw