Operation
Memocheck Sim CYP03D
38
Hauser
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Tau too low
Tau too high
No signal dropout
No amplitude
Temp. too low
Temp. too high
LED voltage
No LED current
Dynamic error
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
Partial pressure
0.0 to 440.0
hPa
0.0 hPa
0.0 to 440.0
hPa
440.0 hPa
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Saturation
-0.02 to
1000 % sat
-0.02 % sat
-0.02 to
1000 % sat
1000 % sat
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Conc. liquid
-0.02 to
120.0 mg/l
-0.02 mg/l
-0.02 to
120.0 mg/l
120.0 mg/l
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to 60.0
˚C
-40.0 °C
-40.0 to 60.0
˚C
60.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Salinity
0.0 to 40.0
g/kg
0.0 g/kg
0.0 to 40.0
g/kg
40.0 g/kg
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to 60.0
˚C
-40.0 °C
-40.0 to 60.0
˚C
60.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Slope
0.0 to
200.0 %
0.0 %
0.0 to
200.0 %
200.0 %
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s