Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the
next pipe elbow: h ≥ 5 × DN
Installation position
Installation position
Recommendation
Vertical orientation
A0017337
Horizontal orientation, transmitter head
up
A0015589
Horizontal orientation, transmitter head
down
A0015590
Horizontal orientation, transmitter head
at side
45°
A0040154
1)
Risk of air pockets.
2)
Risk of build-up of solids.
3)
This installation method in horizontal pipes ensures that the electrodes are always immersed in the
flowing medium.
Alignment with flow direction
The sensor must be installed in such a way that the electrode axis is always at a 90° angle
to the flow direction. The cable glands, which are located on the same axis, serve as a
visual aid.