Float Gauge LTC2230/LTC2240
14
Hauser
4.3.3
Calibration Water Filling Procedure
Indicator adjustment is performed during water filling test. Measure the water level using a measuring tape,
then align indicator to the calculated value. After calculating the value, correct for the value error difference of
submerging float in water and in any liquid other than water (density = 1g/cm3), using the following formula.
After setting the indicator to the measured value for water, determine Lb in the following formula. If the value
of Lb is a positive integer, add that value to the determined indicator value. If the value of Lb is a negative
integer, subtract that value from the determined indicator value. This will be the final indicator value.
4 Installation Adjustment