Entry of exact concentration value for original sample (function 3)
Description
Display
Info
Select function “Corr” in first field of
function group,
confirm with “E”.
Entry of concentration value of
undiluted original sample with “+”
and “–”. Accept the value with “E”.
A concentration value
between 0 and the upper
range value is permissible.
7.7
Calibration result data
After calibration, the characteristics of the current calibration data set are displayed in this
function group and can be read out.
Relative calibration results
The relative calibration results displayed in this function group provide information on the quality
of the previously performed calibration. If the 3-point calibration is good, the difference between
the three relative calibration results for calibration points 1, 2 and 3 should not exceed 50%.
Instrument configuration
liquisys CUM 252
38
Hauser