Liquistation CSF28
Operation
Hauser
37
• Dosing chamber
•
Sample temperature
(for vacuum pump)
• Power failure
1. For the settings to take effect, stop all programs on the start screen beforehand with
the
MODE
soft key.
2. Select
System
/
Sampling
in the menu.
The aforementioned basic settings can now be configured.
Creating sampling programs using the wizard
A Program Wizard is available for the configuration of one or more sampling programs.
The following settings can be made in the Wizard:
• Sampling mode
• Sample volume
• Sampling interval
• Time interval
• Bottle synchronization
• Stop condition
‣
In the
Guidance
menu, start the
Program wizard
and follow the instructions.
A wizard now guides you through the settings.
Performing sampling manually
A sample can be taken manually without a program.
1. Make the desired changes in the
Application
/
Manual sampling
menu or directly
with the
MAN
softkey. This pauses any program currently running.
2. Perform manual sampling with
Start sampling
.
The current bottle configuration and the current sample volume are displayed.
You can select the distributor position. In the case of peristaltic pumps, you can
also change the sample volume.
3. After a manual sampling has been performed, press
ESC
to display and resume an
active program.
The sample volume for "Manual sampling" is not taken into account in the
calculated bottle volumes.
Parameter description
Function
Info
Bottle distribution
Use this function to select the distributor position.
Combination of number of bottles and bottle volume, e.g. 1x60 l (15.85 gal),
4x13 l (3.43 gal).
Can only be configured with the Commissioning Wizard.
Bottle volume
Here it is possible to set the bottle volume to a lower value than the distribution
value set in the Commissioning Wizard, e.g. 4x17 l (4.49 gal) can be reduced to
15 l (4 gal)
Dosing volume
(For version with vacuum
pump)
Sampling volume
(for version with peristaltic
pump)
The dosing volume can be set in the case of the vacuum pump; in the case of
peristaltic pumps, the item is not displayed; the sampling volume is entered
directly in the Program Wizard or can be changed directly in the program in the
Application
/
Program
menu.
Dosing chamber
Only for vacuum pump
Dosing with pressure e.g. in conditions with low suction heights and slight
counterpressure or low volumes.
Sample temperature
Sample temperature setting