Mounting
Liquiphant FTL62 Density
16
Hauser
A0039742
9
Support in the event of dynamic load
5.2
Mounting the device
5.2.1
Required tool
• Open-ended wrench for securing the flange
• Allen key for housing locking screw
5.2.2
Installation
Align the tuning fork using the marking
The tuning fork can be aligned using the marking in such a way that the medium drains
off easily and buildup is avoided.
A0042207
10
Markings to align the tuning fork
Installing in pipes
• Flow velocity up to 5 m/s with a viscosity of 1 mPa⋅s and density of 1 g/cm
3
(SGU).
Check for correct functioning in the event of other process medium conditions.
• Flow velocity > 2 m/s: Separate the tuning fork from the direct flow of media by using
design features such as a bypass or pipe expansion for reduction.
• The flow will not be significantly impeded if the tuning fork is correctly aligned and the
marking is pointing in the direction of flow.
• The marking is visible when installed.