System integration
Liquiline Compact CM72
20
Hauser
Cable
Function
Gray (GY)
Grounding, GND
BU (blue)
4 to 20 mA +
White (WH)
4 to 20 mA -
The grounding cable must be provided by the customer.
5.1
Environment
Degree of pollution of entire device:
Pollution level 4
Degree of internal pollution:
Pollution level 2
Installation category:
Installation category 1
Max. height:
2000 m (6561.68 ft)
Degree of protection:
NEMA Type 6
IP 67
IP 68
Ambient temperature range:
-20°C to 85°C (-4 to 185 F°)
Storage temperature:
-40 ... +85 °C (-40 ... 185 °F)
Relative humidity:
5 ... 95 %
6
System integration
6.1
Integrating the measuring device into the system
Interface for measured value transmission:
4 to 20 mA
For configuration with the measured value and the current output turndown, select the
option in the order structure when ordering. This cannot be changed at a later stage.
Summary of Contents for Liquiline Compact CM72