Proline t-mass I 300 HART
Commissioning
Hauser
99
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface /
Selection / User
entry
Factory setting
Terminal number
–
Shows the terminal numbers
used by the relay output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Relay output function
–
Select the function for the
relay output.
• Closed
• Open
• Diagnostic
behavior
• Limit
• Flow direction
check
• Digital Output
–
Assign limit
The
Limit
option is selected in
the
Relay output function
parameter.
Select process variable for limit
function.
• Temperature
• Mass flow
• Corrected volume
flow
• FAD volume flow
• Volume flow
• Energy flow
• Heat flow
• Density
• Flow velocity
• 2nd temperature
delta heat
• Electronic
temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
–
Assign diagnostic behavior
In the
Relay output function
parameter, the
Diagnostic
behavior
option is selected.
Select diagnostic behavior for
switch output.
• Alarm
• Alarm or warning
• Warning
–
Assign status
In the
Relay output function
parameter, the
Digital Output
option is selected.
Select device status for switch
output.
• Off
• Low flow cut off
–
Switch-off value
In the
Relay output function
parameter, the
Limit
option is
selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
–
Switch-off delay
In the
Relay output function
parameter, the
Limit
option is
selected.
Define delay for the switch-off
of status output.
0.0 to 100.0 s
–
Switch-on value
The
Limit
option is selected in
the
Relay output function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Depends on country
and nominal
diameter
Switch-on delay
In the
Relay output function
parameter, the
Limit
option is
selected.
Define delay for the switch-on
of status output.
0.0 to 100.0 s
–
Failure mode
–
Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
–
*
Visibility depends on order options or device settings
9.4.12 Configuring the local display
The
Display
wizard guides you systematically through all the parameters that can
configured for configuring the local display.