Prosonic M
3 Function group "basic setup" (00)
Hauser
15
add. agitator
fast change
standard solid
Moving surfaces (poss. with
vortex formation) due to agitators
Rapid level change, particularly
in small tanks
For all bulk solids applications
which do not fit in any of the
following groups.
Special filters for stabilising the
input signal are set to large
values.
-> Stable measured value
-> Medium reaction time
The averaging filters are set to
small values.
-> Rapid reaction time
-> Possibly unstable measured
value
The filter and output damping are
set to average values.
solid dusty
conveyor belt
Test: no filter
Dusty bulk solids
Bulk solids with rapid level
change
All the filters can be switched off
for purposes of service and
diagnosis.
The filters are set to detect even
relatively weak signals.
The averaging filters are set to
small values.
-> Rapid reaction time
-> Possibly unstable measured
value
All filters off
HAUSER