![Endress+Hauser FMU 40 Description Of Instrument Functions Download Page 13](http://html.mh-extra.com/html/endress-hauser/fmu-40/fmu-40_description-of-instrument-functions_2407658013.webp)
Prosonic M
3 Function group "basic setup" (00)
Hauser
13
3
Function group "basic setup" (00)
3.1
Function "measured value" (000)
This function displays the current measured value in the selected unit
(see "
customer unit
"
(042)
function). The number of places after decimal point can be
selected in the "
no.of decimals
"
(095)
function.
3.2
Function "tank shape" (002)
This function is used to select the tank shape.
Selection
Þ
Þ
Þ
HAUSER
Prosonic M
dome ceiling
flat ceiling
sphere
horizontal cyl
bypass
stilling well
(ultrasonic guide pipe)
no ceiling
e.g. dumps, open levels
chanels, weirs