3.2
Installation of electrodes and measuring cables
Note:
•
Close off unused electrode
positions with dummy plugs.
•
2 electrodes with KCI fluid and
hose connection cannot be
installed.
•
Before screwing in the electrodes
inspect seat and cleanliness of the
O-ring seal.
•
Inspect all the O-rings on the flow
assembly for damage at regular
intervals.
•
Required mounting clearance
above protection cover: 120 mm
PARAM
MEAS
CAL
DIAG
Insert electrode holder into flow
vessel
Screw in electrode until hand
tight
Connect electrode cable to junction box or to instrument
Secure with union nut
Remove dummy plugs
Tighten another ¼ turn with
socket spanner
1/4
Screw electrode plug onto
electrode head
Attach PM
plug to
PM pin
Attach on protection cover
MONTAGE1.CDR