EngyCal RS33
Wiring
Hauser
25
Pulse output of the flow
sensor
Setting at the
Rx33
Electrical connection
Notes
Active current
A0015363
Pulse I
A
B
50
51
A0015357
A Sensor
B Rx33
The switching
threshold is
between 8 mA
and 13 mA
Namur sensor (as per
EN60947-5-6)
Pulse ID/IE up to
25 Hz or up to
12.5 kHz
10
11
A
B
+
–
A0015359
A Sensor
B Rx33
No monitoring
for short circuit
or line break
takes place.
Voltage pulses and transmitters according to Class IB and IC
(low switching thresholds, small currents)
≤ 1 V corresponds to Low level
≥ 2 V corresponds to High level
U max 30 V, U no-load: 3 to 6 V
Floating contacts,
reed transmitters
Transmitters to Class ID and IE for higher currents and
power supplies
≤ 1.2 mA corresponds to Low level
≥ 2.1 mA corresponds to High level
U no-load: 7 to 9 V
Hauser flowmeters
Flow sensors with PFM or
pulse output:
Proline Prowirl 72 and
Proline Prosonic Flow 92F
Prowirl 72
Prosonic Flow 92F
EngyCal
1 +
2
1 +
A
B
90
90
91
91
50
51
10
11
2
3+
4
A0014181
A = PFM
B = pulse: Terminals 90/91 transmitter power supply, alternatively via external supply
unit