Expert → Diagnosis → Sensor HP → Counter P < Pmin (262)
Write permission
No write permissions. Parameter is read only.
Description
Displays the negative pressure counter for the respective sensor module. The counter is
incremented each time error 841 occurs. You can reset this value using the "Reset Peakhold
(161)" parameter.
COUNTER P > Pmax (263)
Navigation
Expert → Diagnosis → Sensor HP → Counter P > Pmax (263)
Write permission
No write permissions. Parameter is read only.
Description
Displays the overpressure counter for the respective sensor module. The limit value is:
upper sensor module nominal value + 10% of upper sensor module nominal value. You can
reset this value using the "Reset Peakhold (161)" parameter.
MIN. MEAS. TEMP. (264)
Navigation
Expert → Diagnosis → Sensor HP → Min. Meas. Temp. (264)
Write permission
No write permissions. Parameter is read only.
Description
Displays the smallest temperature measured in the sensor module. You can reset this value
using the "Reset Peakhold (161)" parameter.
MAX. MEAS. TEMP. (265)
Navigation
Expert → Diagnosis → Sensor HP → Max. Meas.Temp. (265)
Write permission
No write permissions. Parameter is read only.
Description
Displays the largest temperature measured in the sensor module. You can reset this value
using the "Reset Peakhold (161)" parameter.
Min. meas. press. (266)
Navigation
Expert → Diagnosis → Sensor LP → Minimum Pressure (266)
Write permission
No write permissions. Parameter is read only.