Cleanfit CUA451
Maintenance
Hauser
31
7.
1
1
A0038433
Release the grub screw (item 1) on the underside of the bayonet nut.
8.
1
2
A0038434
Unscrew the bayonet nut and the sensor holder (item 1) from the retraction pipe. In
doing so, hold the retraction pipe steady and turn the handles (2) counterclockwise
(approx. 9 turns).
9. Holding the handles, pull the bayonet nut, along with the sensor holder, out of the
retraction pipe.
10. Unscrew the sensor from the sensor holder.
8.1.3
Cleaning the assembly
To ensure stable and reliable measurements, the assembly and the sensor must be cleaned
regularly. The frequency and intensity of the cleaning process depend on the medium.
A typical example of a cleaning interval would be 6 months in the case of drinking
water.
Cleaning the assembly
1. Remove light dirt and fouling with suitable cleaning solutions.
2. Remove heavy soiling using a soft brush and a suitable cleaning agent.
3. For very persistent dirt, soak the parts in a cleaning solution. Then clean the parts
with a brush.