Cleanfit CUA451
Installation
Hauser
25
For sensors with the Memosens plug-in head
‣
Release the Memosens cable on the sensor.
Changing the position of the locking ring
‣
A0038801
Move the locking ring into the correct groove that corresponds to the sensor type.
The sensor marking on the sensor holder is no longer visible. → 8
Installing the sensor
1.
A0038441
Screw the sensor hand-tight into the internal thread of the sensor holder.
2. Fit the mounted sensor into the retraction pipe.
3. Hold the retraction pipe steady and tighten the bayonet nut (turn the handles
clockwise).
4. Tighten the grub screw of the bayonet nut.
5. Connect the rinse chamber connection.
6. Push up the hand lever as far as possible.
The ball valve is open.
7. Grease the retraction pipe.
This helps the retraction pipe move more smoothly in the direction of the
measuring position.
8. Secure the retraction pipe with the securing screws.