CleanFit CPA 451
Installation
Hauser
9
Orientation
"
Caution!
•
Avoid a siphon effect
1
at the rinse chamber outlet when installing with inclined
orientation. The inlet to the rinse chamber must be from below.
3.3
Installation instructions
3.3.1
Measuring system
A complete measuring system comprises:
•
CleanFit CPA 451 assembly
•
OrbiPac W CPF 81/82 pH/ORP electrode
•
Liquisys M CPM 223/253 transmitter
C07-CPA451xx-14-07-00-xx-001.eps
Fig. 9:
Measuring system
1
OrbiPac W CPF 81/82 pH / Redox electrode
2
Liquisys M CPM 223/253
3
CleanFit CPA 451
4
Medium flow direction
C07-CPA451xx-11-07-00-xx-003.eps
Fig. 7:
Ascending pipe and tank side installation
α
min. 15°
1
Medium flow direction
C07-CPA451xx-11-07-00-xx-002.eps
Fig. 8:
Tank top installation
α
min. 15°
1
Medium flow direction
α
1
90°
α
α
α
1)
Siphon effect: line emptied by vacuum
1
2
3
4