Electrical connection
Cerabar PMP51B Analog
26
Hauser
10.5
11.0
12.5
13.5
14.5
15.5
16.5
17.5
18.5
I
LOOP
U
3.6
3.8
4
4.2
4.4
4.6
4.8
5
5.2
5.4
A
B
A0047149
A
Bluetooth can be switched on
B
Bluetooth cannot be switched on
6.2.2
Terminals
• Supply voltage and internal ground terminal: 0.5 to 2.5 mm
2
(20 to 14 AWG)
• External ground terminal: 0.5 to 4 mm
2
(20 to 12 AWG)
6.2.3
Cable specification
• Protective ground or grounding of the cable shield: rated cross-section > 1 mm
2
(17
AWG)
Rated cross-section of 0.5 mm
2
(20 AWG) to 2.5 mm
2
(13 AWG)
• Cable outer diameter: Ø5 to 12 mm (0.2 to 0.47 in) depends on the cable gland used
(see Technical Information)
6.2.4
4-20 mA
2
3
4
Y
+
+
-
-
I
1
mA
A0028908
2
Block diagram
1
Device
2
Load
3
Power supply
4
multimeter
6.2.5
Overvoltage protection
Devices without optional overvoltage protection
Equipment from Hauser fulfills the requirements of the product standard IEC /
DIN EN 61326-1 (Table 2 Industrial Environment).
Depending on the type of port (DC power supply, input/output port) different testing levels
according to IEC / DIN EN 61326-1 against transient overvoltages (Surge) are applied
(IEC / DIN EN 61000-4-5 Surge):
Test level on DC power ports and input / output ports is 1000 V line to earth
Summary of Contents for Cerabar PMP51B
Page 59: ......