Operation options
Cerabar PMC51B HART
18
Hauser
7
Operation options
7.1
Operating keys and DIP switches on the electronic insert
1
2
Alarm
on
of
f
Max
SW
Zero/
I
Span/
II
Display
3
4
A0039285
1
Operating key for lower range value (Zero)
2
Operating key for upper range value (Span)
3
DIP switch for alarm current
4
DIP switch for locking and unlocking the device
The setting of the DIP switches has priority over the settings made via other operation
methods (e.g. FieldCare/DeviceCare).
7.2
Access to the operating menu via the local display
7.2.1
Device display (optional)
Functions:
Display of measured values and fault and notice messages
The device displays are available with the additional option of Bluetooth® wireless
technology.
Depending on the supply voltage and the current consumption, Bluetooth can optionally
be switched on or off.