Cerabar PMC11, PMC21, PMP11, PMP21
26
Hauser
Environment
Ambient temperature range
Device
Ambient temperature range
PMC11
PMP11
–40 to +70 °C (–40 to +158 °F)
PMC21
PMP21
–40 to +85 °C (–40 to +185 °F)
PMC21
PMP21
Devices for hazardous areas: –40 to +70 °C (–40 to +158 °F)
1)
Exception: the following cable is designed for an ambient temperature range of
–25 to +70 °C (–13 to +158 °F): Product Configurator order code for "Enclosed accessories" option "RZ".
Storage temperature range
–40 to +85 °C (–40 to +185 °F)
Climate class
Device
Climate class
Note
PMC11
PMP11
PMC21
PMP21
Class 3K5
Air temperature: –5 to +45 °C (+23 to +113 °F),
relative humidity: 4 to 95 %
satisfied according to IEC 721-3-3 (condensation not possible)
Degree of protection
Device
Connection
Degree of protection
Option
PMP21
PMP21
Cable5 m (16 ft)
IP66/68
NEMA type 4X/6P enclosure
A
PMP21
PMP21
Cable10 m (33 ft)
IP66/68
NEMA type 4X/6P enclosure
B
PMP21
PMP21
Cable25 m (82 ft)
IP66/68
NEMA type 4X/6P enclosure
C
PMC11
PMP11
M12 plug
IP65 NEMA type 4X enclosure
L
PMC21
PMP21
M12 plug
IP65/67 NEMA type 4X enclosure
M
PMC11
PMP11
PMC21
PMP21
Valve plug ISO4400 M16
IP65 NEMA type 4X enclosure
U
PMC11
PMP11
PMC21
PMP21
Valve plug ISO4400 NPT ½
IP65 NEMA type 4X enclosure
V
1)
Product Configurator order code for "Electrical connection"
2)
IP 68 (1.83m H2O for 24 h)
Vibration resistance
Test standard
Vibration resistance
IEC 60068-2-64:2008
Guaranteed for 5 to 2000Hz: 0.05g
2
/Hz
Electromagnetic
compatibility
• Interference emission as per EN 61326-1 equipment B
• Interference immunity as per EN 61326-1 (industrial environment)
• NAMUR recommendation EMC (NE21)
• Maximum deviation: 1.5% with TD 1:1
For more details please refer to the Declaration of Conformity.