6.2
Version M
Version M is operated via the keys on the
measuring transmitter CPM 152. Please refer
to Operating Instructions BA 143C (order
no. 50077395) for key functions and the com-
plete menu structure of the transmitter.
• In the »Instrument data / system configurati-
on« menu, select the »Chemoclean« clean-
ing function under »Output contacts«.
• Set the parameters for the cleaning functi-
on in the »Instrument data / Chemoclean«
menu. The options are similar to those
available for programming of the CYR 20
(see sections 6-6.1.1).
• Under »assembly« in the »Quick operati-
on« menu, you can move the assembly to
the service position and start a cleaning cy-
cle.
6.2.1
Interval cleaning / weekly program
After expiration of the time interval, a cleaning
cycle is started at time t
0
, then the system
waits for another time interval to expire before
starting the next cleaning cycle. Fixed starting
times are defined for each day of the week in
the case of the weekly program.
6.2.2
Interval measurement
The electrode is outside the medium during
the dwell time and is only moved to the meas-
uring position for the duration of the measur-
ing time. Then the electrode is moved back to
the service position and cleaned.
Hold
Assembly
Drive water
Cleaner
PC20-6.CDR
t
0
start cleaning
t
0
-t
1
3s wait period
t
1
-t
2
assembly movement
t
2
-t
3
pre-rinse time
t
3
-t
4
cleaning time
t
4
-t
5
soak time
t
5
-t
6
post-rinse time
t
6
-t
7
3s wait period
t
7
-t
8
assembly movement
t
8
-t
9
hold continuation time
t
0
t
1
t
2
t
3
t
4
t
5
t
6
t
7
t
8
t
9
Fig. 6.4
Cleaning cycle
sequence
Hold
Assembly
Drive water
Cleaner
t
0
t
1
t
2
t
3
t
4
t
5
t
6
t
7
t
8
t
9
Time t
PC20-7.EPS
t
0
start of measuring
t
0
-t
1
assembly movement
t
1
-t
2
hold continuation time
t
2
-t
3
measuring time
t
3
-t
4
assembly movement
t
4
-t
5
pre-rinse time
t
5
-t
6
cleaning time
t
6
-t
7
soak time
t
7
-t
8
post-rinse time
t
8
-t
9
dwell time
Fig. 6.5
Interval measurement
sequence
06PC20.CHP
autoclean CPC 20 / CPC 20-Z
6
Operation
Hauser
25