• Ensure that the assembly is disconnected from the compressed air supply before you open the
cover.
• Secure the opened assembly against unintentional start-up whilst you carry out work on it.
For cyclic maintenance of the assembly, the shell seals must be replaced. The maintenance interval
(replacement of the shell seals) is dependent on the condition of the medium and must be
determined individually.
Corresponding maintenance instructions are enclosed with the spare parts
• Seal set 50 ml, Order no.
51005590
and
• Seal set 5 ml, Order no.
51004746
8
Accessories
When ordering accessories or spare parts, please state the serial number of the machine.
9
Trouble-shooting
9.1
Trouble-shooting instructions
Always begin trouble-shooting using the following checklists if faults occur after commissioning or
during operation. Different questions will guide you to the cause of the error and will suggest
appropriate remedial action.
9.2
Process error messages
When connecting the ASP Inline sampler assembly to the ASP Station 2000 stationary sampler,
please also observe the information about trouble-shooting in the ASP Station 2000 operating
instructions (
Æ
chapter 9).
Designation
Order code
Control cable for ASP Inline control unit 10x0.22 (connection to the ASP Station 2000
stationary sampler)
51005922
Hose, PVC, clear 12x2
50050376
Pneumatic coupling socket, NW5 with hose clip for LW6 hose
51005589
Error message
Cause
Remedy
ERROR: Dos.up
• Pneumatics defective/
Pneumatics not connected
• Reed contact defective or
incorrectly set
• Pneumatic cylinder defective or
jammed
• Control unit signal incorrect or
not wired
• Check the pneumatic connection
• Check reed contact
Check position of the reed
contacts
E+H Service, if necessary
• Check pneumatic cylinder. E+H
Service, if necessary
• Check wiring
ERROR: Dos.down
see ’ERROR: Dos.up’
ERROR: dosing cylinder
• Reed contact defective or
incorrectly set
• Input defective