Specifications
Biometra TOne
68
10
Specifications
10.1 Model-specific data
Models
Biometra TOne 96 (without gradient function)
Biometra TOne 96 G (with gradient function)
Capacity
96 x 0.2 ml tubes, 96-well micro plates or 8-well strips
Block material
Aluminum (Al) coated with special alloy
Block format
96-well (12 columns)
Number of blocks
1 block, exchanging the block not possible
Sample volumes (recom-
mended)
5 … 50 µl
Sample vessels
Can be used for low-profile and high-profile plastics (plates,
tubes or 8-well strips), plates with and without edge and
with half edge
Temperature range
3 to 99 °C
Control accuracy
±0.1 °C
Maximum heating rate*
4.0 °C/s
Average heating rate*
3.7 °C/s
Maximum cooling rate*
3.3 °C/s
Average cooling rate*
* measured in the block
3.0 °C/s
Temperature uniformity
95 °C
±0.60 °C after 15 s
70 °C
±0.30 °C after 15 s
55 °C
±0.20 °C after 15 s
Gradient (only for Biometra TOne 96 G)
Max./min. Gradient
20 °C/0.1 °C
Gradient temperature range
20 … 99 °C across 12 columns
Heated lid
Heated lid
High-Precision Smart Lid (HPSL)
Lid temperature
30 to 110 °C
Contact pressure
Approximately 10 kg
The clutch mechanism inside the heated lid guarantees a constant
contact pressure on the inserted sample tubes.
10.2 General characteristics
General characteristics
Dimensions (W x D x H)
260 x 430 x 210 mm
Dimensions, with open lid
(B x T x H)
260 x 430 x 385 mm
Mass
11.5 kg
Noise emissions
≤45 dB(A)