Installation and Setup
TOPwave
24
Note:
Pressing the [
LOGGED IN AS
...] key in the Main menu allows to enter login window again
to change login status.
User rights
User
Master
Service
Change user program
No
yes
yes
Change pre-installed applications
No
no
yes
Change type of vessels in program
Yes
yes
yes
Open Setup menu
No
yes
yes
Open Service menu
No
no
yes
Change User Code
No
yes
yes
Change Master Code no
No
yes
yes
Change Service Code
No
no
no
Open the lid by pressing the [O
PEN
] key on the Main menu.
A clearly audible clicking is heard and unlocking is indicated on the screen.
Then open the lid with the red handle. The lid then swings to the right to its end
stop.
After 10 seconds the lid is automatically locked again. In this case, again press [
O-
PEN
].
The rotor for the pressure vessels is preassembled on delivery. Remove the pressure
vessels and insert the empty rotor into the microwave oven. By gently turning the ro-
tor on the axis, find the position at which the rotor again engages downwards.
The rotor is now correctly positioned in the micro-wave oven.
The opening station for CX 100 pressure vessels can be secured in an exhaust hood /
vent unit by means of the adhesives strips on its underside.
5.2.1
Selecting Setup Parameters
Press [S
ETUP
] in M
AIN
menu window.
Note:
Only the user with M
ASTER
login can open the S
ETUP
menu.
The following parameters can be selected:
Turntable installation
Accessories