Commissioning
multi EA 4000
66
5.4.2
Installing the boat sensor
The boat sensor monitors the boat transfer from the sampler FPG 48 to the furnace. It
detects the following fault conditions:
Ceramic boat broken
No ceramic boat on the hook
If one of the fault conditions occurs, the measurement is canceled immediately. An er-
ror message is issued in the multiWin software.
If the boat sensor is not already installed on the autosampler, you can mount it on the
autosampler as follows:
1.
Slide the boat sensor from the left onto
the short guide rail of the FPG 48.
2.
Using the two hexagon socket screws
tighten the boat sensor.
3.
Connect the boat sensor to the "aux"
connection at the rear of the FPG 48.
To resume a measurement after cancellation:
1.
Acknowledge the error message in the software.
2.
If necessary, remove the broken parts of the boat from the furnace and the hook.
3.
Slide the ceramic hook manually to the stop plate.
4.
Reinitialize the device.
5.4.3
Retrofitting the sampler FPG 48 with a boat storage
For measurements using the automatic TIC solids module the sampler must be con-
verted. The following components must be retrofitted:
Boat storage (block with ejection rail for used boats)
Ceramic hook with long sliding rod
Cover for ceramic hook and sliding rod
With the boat storage, the sampler can discard used boats wetted with acid in a waste
container. The use of the boat storage is recommended to prevent unnecessary dam-
age to the boat carousel from acid splashes when returning used boats.
The long sliding rod for the ceramic hook is required to allow the boats to be moved
across the TIC reactor to the furnace of the basic device during TOC determination. The
cover protrudes beyond the sampler on the right-hand side. It prevents the sliding rod
or the ceramic hook from being damaged during the return.
The three components can also be ordered separately.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...