Commissioning
multi EA 4000
48
For the use of the basic device with the furnace module (HTS1500) in areas with
mains voltages of 100 V/115 V/120 V/127 V a 2 phase connection is possible using
optional accessories. Contact Analytik Jena AG if required. This installation may only be
performed by the customer service department of Analytik Jena AG or by specialist per-
sonnel authorized and trained by Analytik Jena AG!
The installation of the electrical equipment of the laboratory must comply with the
standard DIN VDE 0100. After the connection point an electrical current in accordance
with the standard IEC 38 must be available.
5.1.4
Gas supply
The operator is responsible for the gas supply and the corresponding connections and
pressure reducers.
Connection hoses with outer diameter 6 mm and inner diameter 4 mm are included
with the delivery.
Before initial operation, close the flow regulators for argon/oxygen (Ar/O
2
) and oxy-
gen (O
2
) by turning them clockwise (1, 2 in Figure 2 p. 19). Adjust the gas flows later
according to the method used.
5.2
Unpacking and placing the analyzer
WARNING
The analyzer multi EA 4000 must only be set up, assembled and installed by the cus-
tomer service department of Analytik Jena AG or trained specialist personnel author-
ized by Analytik Jena AG!
Any unauthorized intervention in the analyzer can endanger the user and the opera-
tional safety of the equipment and limits or completely invalidates any warranty
claims.
NOTICE
Retain the transport packaging! Return transport for service must be in the original
packaging. This alone prevents transport damage.
The analyzer multi EA 4000 is unpacked and assembled by the customer service de-
partment of Analytik or authorized and trained specialist personnel.
Please check when unpacking the device for completeness and soundness of the deliv-
ery in accordance with the packing list included.
After assembly the customer service tests the analyzer and documents the test.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...