XA02751C
2-Pack/3-Pack TDLAS Gas Analyzer
20
Hauser
To prevent electrostatic discharge
1.
Use a damp cloth to clean the displays to avoid static electricity dis
-
charge.
Replacement Parts
All parts required for operation of the 2-Pack and 3-Pack analyzers must be
supplied by Hauser or an authorized agent. For contact information
to determine specific parts listing for the purchased model → 30.
Fuse Ratings and
Characteristics
Fuses are located on the electronics control board, shown in → 5–1, →
5–2, → 5–3, → 5–4 and → 5–5. Part numbers for the fuse spare part
kits are shown in the Operating Instruction.
Table 5–1 Fuse specifications (Analyzer A)
!
V
A
If you need to replace a fuse, use only the same type and
rating of fuse as the original. Refer also to specifications
listed in Table 5–1 and Table 5–2.
Select the replacement solenoid fuse (F2) based on the
number of solenoids installed on the analyzer.
Drawing
Reference
Voltage
Description
Rating
8
120
VAC
1 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC 0.16A
2 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC 0.25A
3 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC 0.4A
9
Miniature Fuse, 5 x 20 mm,
Time Delay
250 VAC 0.8A
8
240
VAC
1 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.125A
2 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC 0.16A
3 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC 0.2A
9
240
VAC
Miniature Fuse, 5 x 20 mm,
Time Delay
250 VAC 0.8A