XA02751C
2-Pack/3-Pack TDLAS Gas Analyzer
30
Hauser
6.
Restart the SCS.
7.
Check all connections for gas leaks. Using a liquid leak detector is rec
-
ommended.
8.
Re-validate the system with an appropriate gas standard following the
instructions under
“Validating the Analyzer”
in the Description of
Device Parameters for your analyzer.
9.
Purge the scrubber and scrubber efficiency indicator assembly with
nitrogen to remove all flammable gas and cap the inlet and outlet.
Disposal of Used
Scrubbers
1.
Discard used scrubber and scrubber indicator in an appropriate leak-
proof receptacle.
Service
For Service, refer to our website for the list of local sales channels in your
area (https://www.endress.com/contact).
Service Repair Order
If returning the unit is required, obtain a
Service Repair Order (SRO) Num
-
ber
from a Sales channel representative before returning the analyzer to the
factory. Your representative can determine whether the analyzer can be ser
-
viced on site or should be returned to the factory. All returns should be
shipped to:
Hauser
11027 Arrow Rte.
Rancho Cucamonga, CA 91730-4866
United States of America
1-909-948-4100
Depleted H
2
S scrubbers and scrubber indicators contain
predominantly Copper (II) Sulfide [CAS# 1317-40-4] with.
some remaining Copper (II) Oxide [CAS# 1317-38-0] and
basic cupric carbonate [CAS# 12069-69-1], each of which
are odorless dark powders that require few special
precautions other than avoiding contact with the internal
substances, keeping the scrubber tightly sealed and
protecting the contents against humidity.