ENDEVCO 727 Instruction Manual Download Page 5

IM727 Rev A, Page 5 of 9 

 

 

®

ELECTRICAL CONSIDERATIONS

 

 

1.   Excitation Voltage – The drop-test accelerometers are calibrated using an excitation voltage of 

10.000±0.005  Vdc,  unless  otherwise  specified  at  the  time  of  order  (the  maximum  excitation 
voltage  without  damage  is  12  Vdc).  If  a  voltage  other  than  the  voltage  used  at  the  time  of 
calibration is applied to the unit, the ZMO and sensitivity will differ from the specified value on 
the Calibration Certificate, thus excitation voltages other than 10 Vdc should be specified at the 
time of order. The accelerometer requires a clean, well-regulated, constant voltage power supply. 

 

Endevco  offers  a  wide  range  of  signal  conditioners  that  ensure  optimal  performance  from 
Endevco

®

 piezoresistive accelerometers, including the 126, 136 and 4430A. Contact Applications 

Engineering to discuss your specific signal conditioning needs. 

 

2.   Power and Signal Leads – The cable leads are assigned as follows: 

 

 

 

The cable shield is not connected to the accelerometer case. It is recommended that the shield be 
connected to the power supply ground. 

 

3.   Cable  Length  Considerations  –  The  727 accelerometers are  calibrated  with  120  inches  of  cable, 
unless  otherwise  specified  at  the  time  of  order.  When using cables longer than 10 – 20 feet , three  
effects must be taken into account: 

 

a.   

Input Voltage Attenuation

: Resistance in the excitation voltage wires may reduce the 

voltage to the sensor, resulting in a loss of sensitivity. The reduced sensitivity can be 
calculated as follows: 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E

OL  

= loaded output sensitivity 

Eo   = sensitivity into an infinite load (from Cal. Cert.) 

R

O    

= output bridge resistance (from Cal. Cert.), includes cable 

R

L    

= load resistance

Summary of Contents for 727

Page 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Page 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Page 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Page 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Page 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Page 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Page 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Page 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Page 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Reviews: