background image

EN

AEM® Disposable Handpiece and AEM® Disposable Electrodes

Instructions For Use/Care

00282-006   2010/09

1 of 2

   

       

ENCISION recommends placing this document in the Instructions for 

Use/Care Section of your AEM Monitor Operator/Service Manual.

Device Description

Disposable Handpiece - For Single Use Only

The ES1300 series AEM® Disposable Handpiece is intended to be 

used only with Encision’s AEM Disposable Electrodes.

Disposable Electrodes - For Single Use Only

The ES03XX series AEM® Disposable Electrodes are intended to be 

used only with Encision’s ES1300 series AEM Disposable Handle. 

The ES03XX series Electrodes are 5mm instruments designed to fit 

commonly used Trocar Cannulas. The ES03XX series are also sized for 

use with the Disposable Sheath.

Principle of Operation

The Handpiece provides switching for the CUT and COAG functions 

of the ESU. The distal yellow button is CUT. The proximal blue button 

is COAG.

NOTE

• 

This product is rated to 9000 Vp-p. Limit electrosurgical generator 

power setting to 80 Watts or lower (60 Watts for the Conmed 

Aspen Excalibur spray mode). Higher settings may result in 

spurious insulation failure alarms and/or insulation breakdown.  

See Encision AEM Monitor Operator/Service Manual for list of 

compatible electrosurgical generators.

• 

This product is supplied sterile and is not intended for use more 

than one time. No attempt should be made to reprocess this 

device.

Use with Monopolar Electrosurgery

AEM instruments, in conjunction with an AEM Monitor properly 

connected to the electrosurgical generator (ESU), continuously 

monitor and dynamically manage “stray energy” (insulation failure 

and capacitive coupling) in zones 2 & 3, which are likely out of the 

surgeon’s field of view.
AEM shielding does not cover zone 1, which the surgeon should 

keep in view during instrument activation. As in all applications, 

“misapplied” electrosurgical energy remains the responsibility of the 

attending surgeon.

Zone 1

Zone 2

Zone 3

(Note: Zone 1 equals approximately 2.5cm Tip to Shield)

Electrosurgical procedures should be performed only by surgeons 

having adequate training and familiarity with these techniques and 

who are also knowledgeable about anatomy and pathology as well 

as the complications, hazards, risks and benefits of the procedure. 

Indications/Intended Use

The AEM Disposable Handpiece and AEM Disposable Electrodes 

are electrosurgical accessories intended, by use of monopolar 

high-frequency electrical current from compatible electrosurgical 

generators, for ablation, removal, resection and coagulation of soft 

tissue where associated hemostasis is required in open, endoscopic 

and laparoscopic surgical procedures. 
The devices are intended for use by qualified medical personnel 

trained in the use of electrosurgical equipment.
AEM instruments incorporate the use of AEM technology and 

are intended for use with the AEM Monitoring System and 

electrosurgical generators having compatibility with the AEM 

Monitor.

Contraindications

These instruments are not intended for use when electrosurgical 

techniques are contraindicated.

Instructions For Use

Prior to Use

Thoroughly read these instructions and the instructions in the AEM 

Monitor Operator/Service Manual.
The Disposable Handpiece and Disposable Electrodes are supplied 

sterile. Inspect the package and product for damage prior to use.

AEM System Setup

See laminated Setup Sheet (00701) when using the ES9005 series 

AEM Cord Adapter and (02678) when using the ES9015 Universal 

Adapter.

WArNINg

• 

Laparoscopic surgery may result in gas embolism due to 

insufflation of gas into the abdomen.

• 

Pulling or tugging cords can result in breaking of internal 

conductors, causing intermittent alarms or sparking and burning 

of insulation during use. The AEM Monitor is not designed to 

detect or prevent an arc in the event of a broken active wire.

• 

Damage to connector body and/or insulation may result in shock 

or fire hazard.

• 

Keep electrical connections dry while in use to prevent potential 

conduction of HF current to the user.

• 

Damaged external insulation AND incorrect setup of the AEM 

Monitor may result in a risk of unintended patient burn, shock or 

fire hazard. Do not use product having damaged insulation.

CAUTION

• 

Good operating room practice suggests that connections of 

accessories to electrosurgical generators be made only while the 

generator is OFF or on Standby. 

• 

Avoid buildup of eschar on the tip insulation which can provide 

a conductive path for HF energy when in proximity to tissue - by 

either: 

- Use of the Disposable Sheath (preferable), or 

- Frequent cleaning of the tip to remove eschar.

• 

Use these instruments only in conditions that assure adequate 

visualization to minimize risk of misapplied electrosurgical 

energy.

• 

Do not activate the ESU if the monitor’s green Ready light is 

off. No Ready light means there is a setup error or equipment 

malfunction.

• 

Keep ESU power setting as low as possible for the intended 

purpose to minimize unintended burns.

• 

Damaged internal insulation of the cord and/or instrument, 

or loss of shield continuity, may cause ESU return pad alarms 

triggered by the AEM Monitor’s Fault Indicators. For maximum 

patient safety, discontinue use of the instrument if this occurs.

• 

A singular AEM instrument must be the sole conductor of energy 

to tissue. Do not conduct energy by touching an AEM instrument 

to a second instrument contacting tissue. The second device will 

not be protected from capacitive coupling and insulation failure.

• 

Keep electrosurgical instruments away from the patient and 

operative field when not in use. Accidental activation can result in 

unintended injury to the patient.

• 

See electrosurgical generator manual and AEM Monitor 

Operator/Service Manual for precautions concerning the general 

application of electrosurgical equipment.

Assembly Instructions

The ES03XX series AEM Disposable Electrodes are designed to be 

used only with the ES1300 series AEM Disposable Handpiece.
1.  Grasping the rotation knob of the Disposable Electrode, insert it 

into the Handpiece until it snaps or clicks into place. When fully 

seated, the rotation knob should not rotate.

Insert Electrode into Handpiece

Summary of Contents for AEM ES03 Series

Page 1: ...ions These instruments are not intended for use when electrosurgical techniques are contraindicated Instructions For Use Prior to Use Thoroughly read these instructions and the instructions in the AEM Monitor Operator Service Manual The Disposable Handpiece and Disposable Electrodes are supplied sterile Inspect the package and product for damage prior to use AEM System Setup See laminated Setup Sh...

Page 2: ...ion may compromise the integrity of the device which may result in malfunction NOTE Used instruments are considered medical waste Dispose of in accordance with local regulations Express Warranty ENCISION hereby warrants to Buyer that products purchased hereunder shall be free from defects in materials and workmanship under normal use and service as specified in this Instruction for Use Care until ...

Page 3: ...e technieken Instructies voor gebruik Voor gebruik Lees deze instructies en die in de bedienings onderhoudshandleiding van de AEM Monitor grondig door Het handstuk en de elektroden worden steriel geleverd Inspecteer de verpakking en het product voor gebruik op beschadiging Systeeminstelling AEM Raadpleeg het gelamineerde installatieblad 00701 bij gebruik van de AEM snoeradapter uit de ES9005 serie...

Page 4: ...at kan leiden tot storing OPMERKING Gebruikte instrumenten worden beschouwd als medisch afval Afvoeren in overeenstemming met plaatselijke regelgeving Uitdrukkelijke garantie ENCISION garandeert hierbij aan Koper dat hieronder aangeschafte producten vrij zullen zijn van defecten in materialen en vakmanschap bij normaal gebruik en onderhoud zoals gespecificeerd in deze instructies voor gebruik onde...

Reviews: