8 of 44
System Setup (continued)
3. Connect an AEM cord to the EndoShield 2.
4. Connect an AEM Foot Control Instrument to the AEM Cord.
If illuminates GREEN then the EndoShield 2 is fully operational.
If
illuminates RED, see the
Troubleshooting and System Maintenance
section.
If either
or are fl ashing see the
Troubleshooting
section.
5. Turn on the electrosurgical generator, enabling its contact quality
monitoring system.
It should be in its normal operating state.
At End of Life
Removing the Battery
Open the door on the
back of the EndoShield 2
Remove the battery.
The EndoShield 2 contains a 3V CR2 lithium battery. The lithium battery will provide
approximately ten (10) hours of operating time.
WARNING:
Dispose of the lithium battery in accordance with local regulations.
Incineration of the device with battery may result in explosion.
Caution:
The lithium battery is single use only. Do not reuse or replace the lithium
battery.