Troubleshooting and System Maintenance (continued)
Cleaning the EndoShield 2
WARNING:
Electric Shock Hazard. Always unplug the EndoShield 2 before cleaning.
The EndoShield 2 may be cleaned in accordance with your facility's procedures.
1. Disconnect all accessories.
2. Follow the procedures approved by your institution, or use a validated infection
control procedure.
3. Dampen a cloth with a mild cleaning solution (100:1 water to mild detergent, by mass)
or disinfectant and thoroughly wipe all outside surfaces including cords.
4. Ensure unit is dry before use.
Caution:
Do not allow fl uids to enter the EndoShield 2 housing.
Do not clean the EndoShield 2 with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could damage the unit.
Do not steam sterilize the EndoShield 2.
Do not spray the EndoShield 2 directly with cleaning solution.
Storing the EndoShield 2
The EndoShield 2 should be stored within these parameters:
Temperature -13° to 140° F (-25° to 60° C).
Humidity: 5% to 95% relative, non-condensing.
If you store the EndoShield 2 at a temperature that is outside its normal operating range of
59° to 104° F (15° to 40° C), allow one hour for the EndoShield 2 to reach room temperature
before use.