ENCARDIO RITE EDS-71P User Manual Download Page 4

Users’ Manual 

EDS-71V/P Single point borehole extensometer 

  

  

 

Page

 | 

2

 

Dimensional details of sensor 

Displacement 

Overall length of VW sensor 

(in closed position) 

Overall length of Pot. sensor 

(in closed position) 

50 mm 

290 mm 

325 mm 

100 mm 

420 mm

 

425 mm

 

150 mm 

550 mm 

525 mm 

The sensor shaft is fixed to female thread in connecting rod through a sensor lock nut (18), link plate (16), 
sensor tightening nut (17) and a reference button (15) - refer to figures 2.1 and 2.2. The outside body of 
sensor is fixed to a adopter cum guide plate (6) in the flanged housing at the face of the borehole. The 
system is rugged and is protected by a protecting cover (19). Flange assembly (5) .Nipple is an integral 
part of flanged housing (22) through which cable for sensors can be taken to a junction box (23) and then 
through a multi-core cable to any distant location. 

NOTE: 

Encardio-rite uses the convention that depth of anchor is calculated as the distance from the 
mouth of hole to near end of anchor. 

CAUTION: 

Never rotate shaft of displacement sensor in respect to the outside body. This will permanently 
damage the sensor. During installation pull / push or pull shaft only axially. A pin has been 
provided at the end of displacement sensor that sits flush in a groove in sensor body. During 
assembly operation and while using spanners for tightening, keep sensor in closed position 
with pin seated inside groove such that no torque is exerted on its shaft and there is no 
rotational movement. 

1.3   Conventions used in this manual  
WARNING!

  Warning messages calls attention to a procedure or practice, that if not properly followed 

could possibly cause personal injury. 

CAUTION:  

Caution messages calls attention to a procedure or practice, that if not properly followed may 
result in loss of data or damage to equipment. 

NOTE:

 

Note contains important information and is set off from regular text to draw the users’ attention. 

1.4  How to use this manual 

The users’ manual is intended to provide sufficient information for making optimum use of borehole 
extensometers in different applications.  
To make the manual more useful we invite valuable comments and suggestions regarding any   additions 
or enhancements. We also request to please let us know of any errors that are found while going through 
this manual. 

NOTE:

 

The installation personnel must have a background of good installation practices and 
knowledge of the fundamentals of geotechnics. Novices may find it very difficult to carry on 
the installation work. The intricacies involved in installation are such that even if a single 
essential but apparently minor requirement is ignored or overlooked, the most reliable of 
instruments will be rendered useless. 
A lot of effort has gone in preparing this instruction manual. However the best of instruction 
manuals cannot provide for each and every condition in the field, which may affect the 
performance of the instrument. Also, blindly following the instruction manual will not guarantee 
success. Sometimes, depending upon field conditions, the installation personnel will have to 
consciously depart from the written text and use their knowledge and common sense to find 
the solution to a particular problem.      

Summary of Contents for EDS-71P

Page 1: ...ucknow UP 226011 India P 91 522 2661039 42 Email geotech encardio com www encardio com International UAE Qatar Bahrain Bhutan Europe USA India Lucknow Delhi Kolkata Mumbai Chennai Bangalore Hyderabad...

Page 2: ...REHOLE EXTENSOMETER BHE 4 2 1 General Description 4 2 1 1 Reference head assembly 4 2 1 2 Displacement sensor 4 2 1 3 Connecting rod assembly 6 2 1 4 Anchor 7 2 2 Accessories 8 2 2 1 Junction box 8 2...

Page 3: ...ential roof or wall falls before they actually occur Roof or wall falls in underground cavities are almost invariably preceded by measurable sags as the strata opens up and the movement usually occurs...

Page 4: ...his manual WARNING Warning messages calls attention to a procedure or practice that if not properly followed could possibly cause personal injury CAUTION Caution messages calls attention to a procedur...

Page 5: ...he end to get a thorough grasp of the subject You will find a lot of unexpected information in the sections you feel you may skip For description of vibrating wire borehole extensometers manufactured...

Page 6: ...BHE reference head assembly Reference head assembly is supplied packed from factory as per items listed in figure 2 1 The sensor is separately supplied but it is very convenient to assemble it at sit...

Page 7: ...r Junction box assembly Reference button Reference plate Sensor lock nut Sensor tightening nut Link plate Center pillar 17 19 18 21 23 20 Cover 16 15 14 13 22 Nipple with chuck nut Grub screw M4x5 8 b...

Page 8: ...d to downward holes This is because in vertically upward holes the connecting rod is in tension whereas in downward holes it is in compression resulting in sagging or buckling Longer the extensometer...

Page 9: ...in flanged housing its face extends around 20 mm beyond rubber bush 10 Proper sealing in reference head assembly is especially necessary for upwards to horizontally slanting holes for preventing any...

Page 10: ...nsor to the observation room Alternatively a junction box is available for termination of cable from sensor and transmitting the signal through a multi core cable to the observation room 2 2 2 Cable F...

Page 11: ...oth strips soaked in cement during installation 3 1 3 Insert reference head assembly centralised in the 63 mm borehole Taking holes on flange as reference drill four holes 12 mm diameter x 125 mm deep...

Page 12: ...21 4 6821 4 6823 40 05 39 99 50 00 7947 7 7947 7 7946 0 7947 50 00 50 01 Error fs 0 23 0 03 Digit f X 10E 3 Linear gage factor G 8 848E 03 mm digit Thermal factor K 0 003 mm C Polynomial constants A 1...

Page 13: ...9150 1 9150 48 00 0 00 Max error FS 0 16 Linear gage factor G 25 83 mm Volts Displacement D is calculated with the following equation Linear D mm G R1 R0 R1 current reading R0 is initial reading in V...

Page 14: ...le lowest end of grout tube is tied with a cable tie to the connecting rod NOTE In upward to horizontally sloping borehole grout tube should be about 2 m longer than specified depth of anchor and tape...

Page 15: ...ss rod is supplied in coil form using cable ties It is very stiff material and must be carefully handled when unwinding as the end may snap back and cause personal injury 3 3 8 Guide end of fibreglass...

Page 16: ...nsion can be monitored In other words if only compression is expected set zero reading of sensor by extending shaft by around 10 mm If only extension is expected set zero reading of sensor by extendin...

Page 17: ...out pipe to grouting machine or pump grout till the entire cavity is filled and grout flows out of air vent pipe Be very careful that grout flowing out of air vent tube does not get into the flanged h...

Page 18: ...sensor to connecting rod clearly noting position of extensometer 1 CAUTION Never rotate shaft of displacement sensor in respect to the outside body This will permanently damage the sensor During inst...

Page 19: ...nstallation of the multi position borehole extensometer 3 4 1 Soldering iron 25 watt 3 4 2 Rosin 63 37 solder wire 3 4 3 Thread sealant Loctite 290 or equivalent 3 4 4 Cable jointing compound 3 4 5 Ac...