1.6. Enbio PRO sterilization parameters
The table below shows the characteristics of each sterilization program in Enbio
PRO. It is the user's responsibility to select a program suitable for the type of load
and not to exceed the maximum weight of the sterilized load. Failure to observe the
following precautions may compromise patient health and device operation.
* Ambient temperature can have an effect on increasing the total
process time.
* The total duration of the first process may be longer due to the
need for the appliance chamber to warm up.
** After sterilization at 134°C FAST the loads are wet. It is
recommended to leave the drawer ajar in order to evaporate the
remaining water and to lower the temperature of the instruments.
** After sterilization in the 134°C FAST program, the instruments are
intended for immediate use.
Enbio PRO
Program
name
Description of the load
Sterilization
temperature
Sterilization
time
Drying
time
Max. load
weight
Number of
pre-vacuums
Total process
time (for a load of
specified weight)*
Max. water
consumption
Sterilization
process
class
134˚C
solid loads, small porous objects, small porous loads, full porous loads, simple
dimpled objects, narrow clearance objects, multiple packages that may be
unpackaged or packaged (single or multiple layers)
134˚C
4 minutes
4 minutes
800 grams
3
load of 200 grams ≈
18 minutes
190 ml
B
121°C
solid loads, small porous objects, small porous loads, full porous loads, simple
dimpled objects, narrow clearance objects, multiple packages that may be
unpackaged or packaged (single or multiple layers)
121°C
15 minutes
5 minutes
800 grams
3
load of 200 grams ≈
31 minutes
180 ml
B
134 ° C FAST
**
solid, non-porous, simple dental instruments and appliances (e.g. scissors,
handles, pliers, chisels, probes, etc.) only unpackaged, non-textile
134˚C
3.5 minutes
no drying
800 grams
1
load of 100 grams ≈
10 minutes
140 ml
S
134 ° C PRION solid loads, small porous objects, small porous loads, full porous loads, simple
dimpled objects, narrow clearance objects, multiple packages that may be
unpackaged or packaged (single or multiple layers)
134˚C
18 minutes
5 minutes
800 grams
3
load of 800 grams ≈
45 minutes
230 ml
B