3.3.3.5. DC source
Page 3
(Show parameters)
DC POWER SUPPLY
Vb = 13.5 V
-
I = 5.0A
Stop Change
F1
F2
F3
F4
F5
F6
F7
The user can select the parameter to be changed with the related function key and change the value by turning
the knob. Pressing of the ESC button will bring the user back to the previous level from where the test can be
restarted with new parameters.
Page 4
(Change)
DC POWER SUPPLY
Vb = -60.0V - 60.0V
I
= 1A - 10A
Vb
I
13.5
5
F1
F2
F3
F4
F5
F6
F7
Page 5 (Start)
DC POWER SUPPLY
Vb = 13.5 V
- I = 7.0A
I = 5.8 A
Stop Change Zoom
U = 13.4 V
F1
F2
F3
F4
F5
F6
F7
After Start the actual voltage and current values are displayed. All function keys. The latest setting will be
displayed.
Pressing the key F3 while the test is running, the display is switched over to the
ZOOM
mode and is indicating
the actual voltage and current measurement in zoomed letters.
Note: The measured values are for indication only. Low current values
are indicated with the value “LOW”.
The blinking value can be changed with the knob Inc/Dec. To select other values for change use the cursor keys.