Voltage Dips,
short interruptions
and voltage
variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
The
Flex-MT®
Plus
power frequency
magnetic fields should be at levels
characteristic of a typical location in a
typical commercial or hospital environment.
3A/m
NOTE
-UT is the a.c. mains voltage prior to application of the test level.
3A/m
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the
Flex-MT®
Plus
requires continued operation during power
mains interruptions, it is recommended that
the
Flex-MT®
Plus
be powered from an
uninterruptible power supply or a battery.
<5% UT(>95%
dip in UT) for
0.5 cycle
40% UT(60%
dip in UT) for 5
cycles
70% UT(30%
dip in UT) for
25 cycles
<5% UT(>95%
dip in UT) for 5
s
<5% UT(>95%
dip in UT) for
0.5 cycle
40% UT(60%
dip in UT) for 5
cycles
70% UT(30%
dip in UT) for
25 cycles
<5% UT(>95%
dip in UT) for 5
s
Guidance and manufacturer’s declaration-electromagnetic immunity continued
30
Summary of Contents for FLEX-MT PLUS
Page 3: ...CAUTION Federal law requires a prescription from your physician before use of this product ...
Page 11: ...LCD Display 11 Recharger Receptacle ...
Page 34: ...34 2 N Normal Symmetric Single Pulse Waveform Multi Pulse Waveform Asymmetric ...
Page 38: ...38 Symmetric Single Pulse Waveform Channel 2 Multi Pulse Waveform Channel 2 Asymmetric ...
Page 40: ...S Synchronous Symmetric Single Pulse Waveform Multi Pulse Waveform Asymmetric 40 ...