Symptom
Solution
The unit does not start or stop
•
Check the electrical connections and power supply at the mains
•
Check the fuses on back panel
•
Check main switch function: at least one LED must be visible on adjustments and working mode areas
•
The unit is in "standby" mode, tap on any sensitive zone to exit standby
•
Send the unit to an authorized EMS Repair Center
The foot pedal does not activate the unit
• Check that the foot pedal is properly connected
• Replace the foot pedal
•
Send the unit to an authorized EMS Repair Center
The unit is still activated when the foot
pedal is released
•
The foot pedal is active, check for jam at foot pedal
•
The switch in the pedal is blocked. Turn the unit off at the power switch
• Replace the foot pedal
•
Send the unit to an authorized EMS Repair Center
User interface displays an error: there
is one acoustic signal and all LEDs are
blinking
•
Powder chamber pressure is too low or too high
•
The air or liquid regulator is switched off
•
Foot pedal and purge commands are not recognized by system. Check connections
The handpiece does not remain on the
magnetic support
•
Install the magnetic support on a metallic part of the unit
•
The support is demagnetized
•
Replace magnetic support
The handpiece does not hold on the
handpiece cord
• Check for presence of the O-ring seal and replace if necessary
• Replace the handpiece
• Replace the handpiece cord
•
Send the handpiece and the handpiece cord to an authorized EMS Repair Center
No water jet coming from the handpiece
•
Check the water supply and water filter
•
Check the water adjustment of your unit
•
Check the connections between the unit, the cord and the handpiece
•
Check for presence of water at the cord outlet port
•
Send the handpiece to an authorized EMS Repair Center
No powder/air jet coming from the
handpiece
•
Check the quantity of powder in your unit
•
Check the compressed air supply and air filter
•
Check for presence of powder/air at cord outlet port
•
Eliminate residual powder present in the channels using the cleaning needles provided
•
Perio powder chamber only: remove the diffuser and empty the powder. Blow the diffuser with compressed air before
installing it again and refilling.
•
Send the system to an authorized EMS Repair Center
TROUBLESHOOTING
Cont. next page
FB-580_CN_rev_B_01_Not Util AIR-FLOW MASTER.indd 80
13.10.2017 10:00:48
Summary of Contents for AIR-FLOW MASTER Perio
Page 9: ...FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 9 13 10 2017 09 57 34...
Page 15: ...3 2 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 3 13 10 2017 09 57 41...
Page 17: ...5 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 5 13 10 2017 09 57 47...
Page 19: ...7 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 7 13 10 2017 09 57 49...
Page 25: ...13 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 13 13 10 2017 09 58 11...
Page 39: ...27 EMS O Perio PERIO FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 27 13 10 2017 09 59 23...
Page 53: ...41 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 41 13 10 2017 09 59 41...
Page 55: ...43 AIR FLOW FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 43 13 10 2017 09 59 45...
Page 61: ...49 OFF FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 49 13 10 2017 09 59 58...
Page 63: ...51 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 51 13 10 2017 10 00 03...
Page 65: ...53 CLICK FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 53 13 10 2017 10 00 08...
Page 67: ...55 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 55 13 10 2017 10 00 13...
Page 69: ...57 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 57 13 10 2017 10 00 15...
Page 75: ...63 O O FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 63 13 10 2017 10 00 27...
Page 77: ...65 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 65 13 10 2017 10 00 29...
Page 79: ...67 EMS FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 67 13 10 2017 10 00 34...
Page 85: ...73 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 73 13 10 2017 10 00 44...
Page 91: ...79 FB 580_CN_rev_B_01_Not Util AIR FLOW MASTER indd 79 13 10 2017 10 00 48...