Quick Select Back, Low Back/Hip Programs – Modulation Amplitude (MA)
In the Modulation Amplitude program, for 0.5 seconds the intensity is at 100% of the set
level and for the next 0.5 seconds, the intensity is at the set span percentage (60%) of the
preset level. The cycle is then repeated. There is no off time.
Continuous (C)
This setting produces a continuous stimulation at the set intensity. Channel 1 pulses
alternate with Channel 2 pulses.
Quick Select Shoulder, Hand/Wrist Programs –
Simple Modulated Pulse (SMP)
The Simple Modulated Pulse (SMP) program delivers a group of pulses as a repeating
12-second cycle. Within each cycle, the rate and duration of the pulses vary while the
intensity remains constant. This mode is unique because the rate stays in the 2 to 10 pps
range for 1/3 of the cycle time (4 seconds) as the rate modulates down to 2 pulses per
second (pps) and then back up again.
Burst
The Burst program gradually increases the intensity from 0 to the set level over 0.4
seconds, maintains it at the set level for 1.0 seconds, and then gradually decreases the
intensity over 0.5 seconds. The device then sends no stimulation for 3.0 seconds.
Channel 1 pulses alternate with pulses of Channel 2.
Modulation
Rate is selected pseudo-randomly from 8 rates: 2, 10, 20, 40, 60, 80, 100, or 150 pps.
Span percentage for pulse duration is selected pseudo-randomly from 7 values between
50% of set pulse duration and the set pulse duration. Each combination is on for the
selected cycle time.