background image

TB-6520

 Page 4 of 7

Test Lab Test Procedure Guideline

For test lab use of ESD Worksurfaces, Floor Materials, 

Footwear, Garments, or Seating, best advice is to follow 
procedures in applicable ESD Association standards (see 
above) which include details regarding:

•  Cleaning (For example per S4.1 “The test specimens and 

electrodes shall be cleaned twice with a minimum 70% 

isopropanol-water solution using a clean, low-linting cloth 

each time). Allow to dry.

•  Environmental chamber (For example per S4.1 control 

relative humidity to 12 ±3% RH and 50 ±5% RH and 

temperature to 23 ±1 degrees C)

•  Specimen support surface (For example per S4.1 

greater than 1.0 x 10E12 ohms such as PMMA, PTFE or 

polycarbonate)

•  Specimen Pre-Conditioning (For example per S4.1 at 23 

±1 degrees C; 3 specimens at 12 ±3% relative humidity 

for 72 hours minimum, and 3 specimens at 50 ±5% 

relative humidity for 72 hours minimum)

•  Reporting Test Results, for example per S4.1 Reporting 

Test Results, Report:

•  Minimum, median and maximum readings for both  

resistance-to-groundable point and point-to-point 

resistance in ohms at low relative humidity

•  Minimum, median and maximum readings for both  

resistance-to-groundable point and point-to-point    

resistance in ohms at moderate relative humidity

•  Temperature

•  Relative humidity

•  Actual duration of conditioning

•  Test equipment used

Periodic Checks of Installed Product Test 
Procedure Guideline

The ESD Association lists test procedures and 

troubleshooting tips in Compliance Verification ESD TR53.

We recommend this document be purchased from 

www.ESDA.org

Measure RTG Resistance to Ground 

Test Procedure in reference with ANSI/ESD S4.1 section 

6.4 Periodic Worksurface Testing:

•  Do not clean the surface

•  Remove from the surface only those items that might 

interfere with the test. 

•  ESD sensitive devices shall also be removed 

•  The black sensing test lead with shielded plug should go 

to equipment ground or a groundable point

•  Use one Electrode on other test lead and place Electrode 

the farthest convenient point on the surface

•  Press button and hold Test Button until measurement is 

displayed

•  Perform additional measurements placing Electrode on 

the most commonly used or most worn area

If measurement is outside acceptable limits, clean surface 
and retest to determine if cause of failure is insulative dirt 

layer or the ESD contol item.  Note: Use an ESD cleaner 

containing no insulative silicone.

Measure RTT Resistance Between Two Points 
on the Surface

•  Do not clean the surface

•  Remove from the surface only those items that might 

interfere with the test. 

•  ESD sensitive devices shall also be removed 

•  Use two 5 Pound Electrodes, place in the most commonly 

used portion of the surface 10" apart (2" from any edge, 

3" from any groundable point) for worksurface and 3' for 

flooring

•  Press Test button and hold Button until measurement is 

displayed

•  If the most used portion is not obvious, use two points 

near the center of the surface

If measurement is outside acceptable limits, clean surface 
and retest to determine if cause of failure is insulative dirt 

layer or the ESD contol item.  Note: Use an ESD cleaner 

containing no insulative silicone.

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Figure 3. Setting up for RTG testing.

10”

Figure 4. Setting up for RTT testing.  (Electrodes are to be 
about 10" apart for worksurfaces and about 3' for flooring)

Reporting and Using Test Results

Different standards have different requirements; use 

requirements as specified in user’s ESD control plan.  

Examples are:

Per ANSI/ESD 4.1 (worksurfaces), Report:

•  RTG maximum and minimum values measured for 

resistance-to-ground

•  RTT maximum and minimum values measured for 

point-to-point resistance in ohms

EMIT

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Fax (909) 627-7449 • Website: 

DescoEMIT.com

Summary of Contents for 50557

Page 1: ...ting Resistive Measurement Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versati...

Page 2: ...rees Celsius tolerance 3 C typical Humidity as percentage from 5 95 tolerance 10 typical Surface resistance mantissa with exponent displayed via illuminated LED measurement in ohms Measurement being d...

Page 3: ...S4 1 Worksurfaces The test specimens and electrodes shall be cleaned twice with a minimum 70 isopropanol water solution using a clean low linting cloth each time Note then conditioned for 72 hours min...

Page 4: ...ensitive devices shall also be removed The black sensing test lead with shielded plug should go to equipment ground or a groundable point Use one Electrode on other test lead and place Electrode the f...

Page 5: ...hms RTT Surface Resistivity should measure 10E8 ohms square Ref ANSI ESD S11 11 section 12 0 Conversion to Resistivity states When it is appropriate to convert a resistance obtained by this test metho...

Page 6: ...ses will act like an insulator to the Meter Meter will have to be stationary in testing area for about 2 hours with no significant temperature changes CALIBRATION TEST PROCEDURE When testing use only...

Page 7: ...f your original packing slip invoice or other proof of date of purchase Any unit under warranty should be shipped prepaid to the EMIT factory Warranty replacements will take approximately two weeks If...

Reviews: