3
Rosemount Series 58C sheath cutting
Procedure
1. Determine the length to which the sheath will be cut. The finished
length needs to include an additional 1.5-in. (3.8 cm) for
compression fittings or 2.5-in. (6.5 cm) for spring-loaded fittings (see
).
2. Remove and save the heat shrink tubing from the rear of the sensor.
3. Place the sensor in a vise, taking care not to overtighten, and position
the tubing cutter on the sheath.
4. Score the sheath to a depth of approximately 1/64-in. (0.4 cm) To
prevent damage to the lead wire insulation, do not cut completely
through the sheath.
5. Firmly grasp the end of the sheath with your hand or a pair of pliers.
Using a sharp snapping motion, break off and remove the excess
sheath material. Take care not to strip or damage the lead wire
insulation while removing the excess sheath material.
Note
If you are unable to easily break off excess sheath material, deepen
the score and repeat
6. Replace the heat shrink tubing.